REV. 7-10U S E R ’ S G U I D EAUTOTUNE SERIESHIGH INTENSITY ULTRASONIC PROCESSORMICROPROCESSOR CONTROLLED1500 WATT MODEL WITH HIGH VOLUME FLOW CELLTAB
7Press the I key. The screen will display the power rating and the frequency of theUltrasonic Processor, and the following control parameters.AMPLITUD
81. Using the fittings on top of the converter, circulate dry compressed air through theconverter to cool the converter.2. Depress the START key.3. Wi
94. Using the ▲ or ▼ keys, increase or decrease the amplitude as required.5. Process the material at different flow rates, while maintaining the press
PULSER:* The pulse function is not used when working with the High Volume FlowCell, but can be used when processing a batch with a probe. By inhibitin
11TEMPERATURE:* The temperature function prevents overheating of the sample bycontinuously monitoring the sample temperature, and terminating the ultr
12REVIEW: The REVIEW function provides a “window” on the process by displayingvarious operating parameters without process interruption. Pressing theE
13ENERGY SETPOINT:* The energy setpoint continuously monitors the amount ofenergy in Joules (watts-seconds), that is being delivered to the probe, and
14RECALL: The recall function can retrieve from memory any of 9 stored controlparameters for verification or usage. To retrieve any parameters press t
15SECTION III – SERVICE INFORMATIONYour Ultrasonic Processor was designed to provide you with years of safe anddependable service. Nevertheless, becau
16RETURN OF EQUIPMENTIt is suggested that if a unit is in need of repair, both the Ultrasonic Processor and theHigh Volume Flow Cell be sent back to t
WARRANTYYour Ultrasonic Processor is warranted for a period of three years from thedate of shipment against defects in material and workmanship under
17SONICS SAFETY CERTIFICATION FORMItems being returned:Please check only one item below:___ The equipment was never used or exposed to any radiologica
IMPORTANT SAFEGUARDSREAD BEFORE INSTALLING ORUSING THE EQUIPMENTYour Ultrasonic Processor has been designed with safety in mind. However, no designcan
1SECTION I – INSTALLATIONINSPECTIONPrior to installing the Ultrasonic Processor and the High Volume Flow Cell, perform avisual inspection to detect an
2SECTION II – OPERATIONPRINCIPLES OF ULTRASONIC DISRUPTIONThe ultrasonic power supply converts 50/60 Hz line voltage to high frequencyelectrical energ
3Displays prompts and the following control parameters:• Amplitude selected• Output power delivered to the probe in watts, and as percentage of the to
4FUNCTIONS OF KEYS, CONTROLS, INDICATORS, AND CONNECTORS(cont.)REAR PANELTemperature probe jackFootswitch jackConnects to the footswitch cable.Connect
5PREPARATION FOR USE1. Plug the electrical line cord into the electrical outlet. If the unit is already on, switchthe unit off by depressing the O key
6USING THE ULTRASONIC PROCESSORThe speed control on an automobile, can, to a certain extent, be compared to anUltrasonic Processor. The speed control